Along with normal patient management, the following patient management and assessment is recommended for a patient using The Insides® Neo.
Rate of Refeeding
It is recommended a syringe infusion pump is used to reinfuse the patient’s chyme via The Insides® Neo. This ensures the reinfusion is controlled and comfortable for the patient. Each patient will have their own tolerance for the rate of refeeding. As a general guide:
Monitoring Bowel Movements
Patients will likely have a bowel movement within the first three days of starting chyme reinfusion with The Insides® Neo. If a bowel movement has not happened within three days, urgent assessment is required.
Ostomy Appliance
The Insides® Neo couples with a Hollister™ Newborn Pouch “Pouchkins™” 3778 and Coloplast® “Sensura® Mio Baby”. This ostomy appliance must be changed when it leaks and The Insides® Neo must be changed at the same time. The Insides® Neo can be used for up to four days.
Assessment of Pain and Discomfort
Regular assessment of abdominal distention and discomfort is necessary. This should be completed a minimum of every 4 – 6 hours.
Device Troubleshooting
Please refer to the Frequently Asked Questions for troubleshooting The Insides® Neo
‘Coloplast’, ‘SenSura® Mio ’ are registered trademarks of Coloplast Corporation. The use of the trademarks ‘Coloplast’, ‘SenSura® Mio ‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Coloplast Corporation. We hereby affirm our recognition of Coloplast Corporation 's exclusive intellectual property rights in the trademark ‘Coloplast’, ‘SenSura® Mio,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
‘Hollister’, ‘Pouchkins™’ are registered trademarks of Hollister Incorporated. The use of the trademarks ‘Hollister’, ‘Pouchkins™‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Hollister Incorporated. We hereby affirm our recognition of Hollister Incorporated 's exclusive intellectual property rights in the trademark ‘Hollister’, ‘Pouchkins™,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
Along with normal patient management, the following patient management and assessment is recommended for a patient using The Insides® Neo.
Rate of Refeeding
It is recommended a syringe infusion pump is used to reinfuse the patient’s chyme via The Insides® Neo. This ensures the reinfusion is controlled and comfortable for the patient. Each patient will have their own tolerance for the rate of refeeding. As a general guide:
Monitoring Bowel Movements
Patients will likely have a bowel movement within the first three days of starting chyme reinfusion with The Insides® Neo. If a bowel movement has not happened within three days, urgent assessment is required.
Ostomy Appliance
The Insides® Neo couples with a Hollister™ Newborn Pouch “Pouchkins™” 3778 and Coloplast® “Sensura® Mio Baby”. This ostomy appliance must be changed when it leaks and The Insides® Neo must be changed at the same time. The Insides® Neo can be used for up to four days.
Assessment of Pain and Discomfort
Regular assessment of abdominal distention and discomfort is necessary. This should be completed a minimum of every 4 – 6 hours.
Device Troubleshooting
Please refer to the Frequently Asked Questions for troubleshooting The Insides® Neo
‘Coloplast’, ‘SenSura® Mio ’ are registered trademarks of Coloplast Corporation. The use of the trademarks ‘Coloplast’, ‘SenSura® Mio ‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Coloplast Corporation. We hereby affirm our recognition of Coloplast Corporation 's exclusive intellectual property rights in the trademark ‘Coloplast’, ‘SenSura® Mio,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
‘Hollister’, ‘Pouchkins™’ are registered trademarks of Hollister Incorporated. The use of the trademarks ‘Hollister’, ‘Pouchkins™‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Hollister Incorporated. We hereby affirm our recognition of Hollister Incorporated 's exclusive intellectual property rights in the trademark ‘Hollister’, ‘Pouchkins™,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
Along with normal patient management, the following patient management and assessment is recommended for a patient using The Insides® Neo.
Rate of Refeeding
It is recommended a syringe infusion pump is used to reinfuse the patient’s chyme via The Insides® Neo. This ensures the reinfusion is controlled and comfortable for the patient. Each patient will have their own tolerance for the rate of refeeding. As a general guide:
Monitoring Bowel Movements
Patients will likely have a bowel movement within the first three days of starting chyme reinfusion with The Insides® Neo. If a bowel movement has not happened within three days, urgent assessment is required.
Ostomy Appliance
The Insides® Neo couples with a Hollister™ Newborn Pouch “Pouchkins™” 3778 and Coloplast® “Sensura® Mio Baby”. This ostomy appliance must be changed when it leaks and The Insides® Neo must be changed at the same time. The Insides® Neo can be used for up to four days.
Assessment of Pain and Discomfort
Regular assessment of abdominal distention and discomfort is necessary. This should be completed a minimum of every 4 – 6 hours.
Device Troubleshooting
Please refer to the Frequently Asked Questions for troubleshooting The Insides® Neo
‘Coloplast’, ‘SenSura® Mio ’ are registered trademarks of Coloplast Corporation. The use of the trademarks ‘Coloplast’, ‘SenSura® Mio ‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Coloplast Corporation. We hereby affirm our recognition of Coloplast Corporation 's exclusive intellectual property rights in the trademark ‘Coloplast’, ‘SenSura® Mio,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
‘Hollister’, ‘Pouchkins™’ are registered trademarks of Hollister Incorporated. The use of the trademarks ‘Hollister’, ‘Pouchkins™‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Hollister Incorporated. We hereby affirm our recognition of Hollister Incorporated 's exclusive intellectual property rights in the trademark ‘Hollister’, ‘Pouchkins™,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
Along with normal patient management, the following patient management and assessment is recommended for a patient using The Insides® Neo.
Rate of Refeeding
It is recommended a syringe infusion pump is used to reinfuse the patient’s chyme via The Insides® Neo. This ensures the reinfusion is controlled and comfortable for the patient. Each patient will have their own tolerance for the rate of refeeding. As a general guide:
Monitoring Bowel Movements
Patients will likely have a bowel movement within the first three days of starting chyme reinfusion with The Insides® Neo. If a bowel movement has not happened within three days, urgent assessment is required.
Ostomy Appliance
The Insides® Neo couples with a Hollister™ Newborn Pouch “Pouchkins™” 3778 and Coloplast® “Sensura® Mio Baby”. This ostomy appliance must be changed when it leaks and The Insides® Neo must be changed at the same time. The Insides® Neo can be used for up to four days.
Assessment of Pain and Discomfort
Regular assessment of abdominal distention and discomfort is necessary. This should be completed a minimum of every 4 – 6 hours.
Device Troubleshooting
Please refer to the Frequently Asked Questions for troubleshooting The Insides® Neo
‘Coloplast’, ‘SenSura® Mio ’ are registered trademarks of Coloplast Corporation. The use of the trademarks ‘Coloplast’, ‘SenSura® Mio ‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Coloplast Corporation. We hereby affirm our recognition of Coloplast Corporation 's exclusive intellectual property rights in the trademark ‘Coloplast’, ‘SenSura® Mio,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
‘Hollister’, ‘Pouchkins™’ are registered trademarks of Hollister Incorporated. The use of the trademarks ‘Hollister’, ‘Pouchkins™‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Hollister Incorporated. We hereby affirm our recognition of Hollister Incorporated 's exclusive intellectual property rights in the trademark ‘Hollister’, ‘Pouchkins™,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
Along with normal patient management, the following patient management and assessment is recommended for a patient using The Insides® Neo.
Rate of Refeeding
It is recommended a syringe infusion pump is used to reinfuse the patient’s chyme via The Insides® Neo. This ensures the reinfusion is controlled and comfortable for the patient. Each patient will have their own tolerance for the rate of refeeding. As a general guide:
Monitoring Bowel Movements
Patients will likely have a bowel movement within the first three days of starting chyme reinfusion with The Insides® Neo. If a bowel movement has not happened within three days, urgent assessment is required.
Ostomy Appliance
The Insides® Neo couples with a Hollister™ Newborn Pouch “Pouchkins™” 3778 and Coloplast® “Sensura® Mio Baby”. This ostomy appliance must be changed when it leaks and The Insides® Neo must be changed at the same time. The Insides® Neo can be used for up to four days.
Assessment of Pain and Discomfort
Regular assessment of abdominal distention and discomfort is necessary. This should be completed a minimum of every 4 – 6 hours.
Device Troubleshooting
Please refer to the Frequently Asked Questions for troubleshooting The Insides® Neo
‘Coloplast’, ‘SenSura® Mio ’ are registered trademarks of Coloplast Corporation. The use of the trademarks ‘Coloplast’, ‘SenSura® Mio ‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Coloplast Corporation. We hereby affirm our recognition of Coloplast Corporation 's exclusive intellectual property rights in the trademark ‘Coloplast’, ‘SenSura® Mio,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
‘Hollister’, ‘Pouchkins™’ are registered trademarks of Hollister Incorporated. The use of the trademarks ‘Hollister’, ‘Pouchkins™‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Hollister Incorporated. We hereby affirm our recognition of Hollister Incorporated 's exclusive intellectual property rights in the trademark ‘Hollister’, ‘Pouchkins™,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
Along with normal patient management, the following patient management and assessment is recommended for a patient using The Insides® Neo.
Rate of Refeeding
It is recommended a syringe infusion pump is used to reinfuse the patient’s chyme via The Insides® Neo. This ensures the reinfusion is controlled and comfortable for the patient. Each patient will have their own tolerance for the rate of refeeding. As a general guide:
Monitoring Bowel Movements
Patients will likely have a bowel movement within the first three days of starting chyme reinfusion with The Insides® Neo. If a bowel movement has not happened within three days, urgent assessment is required.
Ostomy Appliance
The Insides® Neo couples with a Hollister™ Newborn Pouch “Pouchkins™” 3778 and Coloplast® “Sensura® Mio Baby”. This ostomy appliance must be changed when it leaks and The Insides® Neo must be changed at the same time. The Insides® Neo can be used for up to four days.
Assessment of Pain and Discomfort
Regular assessment of abdominal distention and discomfort is necessary. This should be completed a minimum of every 4 – 6 hours.
Device Troubleshooting
Please refer to the Frequently Asked Questions for troubleshooting The Insides® Neo
‘Coloplast’, ‘SenSura® Mio ’ are registered trademarks of Coloplast Corporation. The use of the trademarks ‘Coloplast’, ‘SenSura® Mio ‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Coloplast Corporation. We hereby affirm our recognition of Coloplast Corporation 's exclusive intellectual property rights in the trademark ‘Coloplast’, ‘SenSura® Mio,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
‘Hollister’, ‘Pouchkins™’ are registered trademarks of Hollister Incorporated. The use of the trademarks ‘Hollister’, ‘Pouchkins™‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Hollister Incorporated. We hereby affirm our recognition of Hollister Incorporated 's exclusive intellectual property rights in the trademark ‘Hollister’, ‘Pouchkins™,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
Along with normal patient management, the following patient management and assessment is recommended for a patient using The Insides® Neo.
Rate of Refeeding
It is recommended a syringe infusion pump is used to reinfuse the patient’s chyme via The Insides® Neo. This ensures the reinfusion is controlled and comfortable for the patient. Each patient will have their own tolerance for the rate of refeeding. As a general guide:
Monitoring Bowel Movements
Patients will likely have a bowel movement within the first three days of starting chyme reinfusion with The Insides® Neo. If a bowel movement has not happened within three days, urgent assessment is required.
Ostomy Appliance
The Insides® Neo couples with a Hollister™ Newborn Pouch “Pouchkins™” 3778 and Coloplast® “Sensura® Mio Baby”. This ostomy appliance must be changed when it leaks and The Insides® Neo must be changed at the same time. The Insides® Neo can be used for up to four days.
Assessment of Pain and Discomfort
Regular assessment of abdominal distention and discomfort is necessary. This should be completed a minimum of every 4 – 6 hours.
Device Troubleshooting
Please refer to the Frequently Asked Questions for troubleshooting The Insides® Neo
‘Coloplast’, ‘SenSura® Mio ’ are registered trademarks of Coloplast Corporation. The use of the trademarks ‘Coloplast’, ‘SenSura® Mio ‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Coloplast Corporation. We hereby affirm our recognition of Coloplast Corporation 's exclusive intellectual property rights in the trademark ‘Coloplast’, ‘SenSura® Mio,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
‘Hollister’, ‘Pouchkins™’ are registered trademarks of Hollister Incorporated. The use of the trademarks ‘Hollister’, ‘Pouchkins™‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Hollister Incorporated. We hereby affirm our recognition of Hollister Incorporated 's exclusive intellectual property rights in the trademark ‘Hollister’, ‘Pouchkins™,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
Along with normal patient management, the following patient management and assessment is recommended for a patient using The Insides® Neo.
Rate of Refeeding
It is recommended a syringe infusion pump is used to reinfuse the patient’s chyme via The Insides® Neo. This ensures the reinfusion is controlled and comfortable for the patient. Each patient will have their own tolerance for the rate of refeeding. As a general guide:
Monitoring Bowel Movements
Patients will likely have a bowel movement within the first three days of starting chyme reinfusion with The Insides® Neo. If a bowel movement has not happened within three days, urgent assessment is required.
Ostomy Appliance
The Insides® Neo couples with a Hollister™ Newborn Pouch “Pouchkins™” 3778 and Coloplast® “Sensura® Mio Baby”. This ostomy appliance must be changed when it leaks and The Insides® Neo must be changed at the same time. The Insides® Neo can be used for up to four days.
Assessment of Pain and Discomfort
Regular assessment of abdominal distention and discomfort is necessary. This should be completed a minimum of every 4 – 6 hours.
Device Troubleshooting
Please refer to the Frequently Asked Questions for troubleshooting The Insides® Neo
‘Coloplast’, ‘SenSura® Mio ’ are registered trademarks of Coloplast Corporation. The use of the trademarks ‘Coloplast’, ‘SenSura® Mio ‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Coloplast Corporation. We hereby affirm our recognition of Coloplast Corporation 's exclusive intellectual property rights in the trademark ‘Coloplast’, ‘SenSura® Mio,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
‘Hollister’, ‘Pouchkins™’ are registered trademarks of Hollister Incorporated. The use of the trademarks ‘Hollister’, ‘Pouchkins™‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Hollister Incorporated. We hereby affirm our recognition of Hollister Incorporated 's exclusive intellectual property rights in the trademark ‘Hollister’, ‘Pouchkins™,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
Along with normal patient management, the following patient management and assessment is recommended for a patient using The Insides® Neo.
Rate of Refeeding
It is recommended a syringe infusion pump is used to reinfuse the patient’s chyme via The Insides® Neo. This ensures the reinfusion is controlled and comfortable for the patient. Each patient will have their own tolerance for the rate of refeeding. As a general guide:
Monitoring Bowel Movements
Patients will likely have a bowel movement within the first three days of starting chyme reinfusion with The Insides® Neo. If a bowel movement has not happened within three days, urgent assessment is required.
Ostomy Appliance
The Insides® Neo couples with a Hollister™ Newborn Pouch “Pouchkins™” 3778 and Coloplast® “Sensura® Mio Baby”. This ostomy appliance must be changed when it leaks and The Insides® Neo must be changed at the same time. The Insides® Neo can be used for up to four days.
Assessment of Pain and Discomfort
Regular assessment of abdominal distention and discomfort is necessary. This should be completed a minimum of every 4 – 6 hours.
Device Troubleshooting
Please refer to the Frequently Asked Questions for troubleshooting The Insides® Neo
‘Coloplast’, ‘SenSura® Mio ’ are registered trademarks of Coloplast Corporation. The use of the trademarks ‘Coloplast’, ‘SenSura® Mio ‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Coloplast Corporation. We hereby affirm our recognition of Coloplast Corporation 's exclusive intellectual property rights in the trademark ‘Coloplast’, ‘SenSura® Mio,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
‘Hollister’, ‘Pouchkins™’ are registered trademarks of Hollister Incorporated. The use of the trademarks ‘Hollister’, ‘Pouchkins™‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Hollister Incorporated. We hereby affirm our recognition of Hollister Incorporated 's exclusive intellectual property rights in the trademark ‘Hollister’, ‘Pouchkins™,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
Along with normal patient management, the following patient management and assessment is recommended for a patient using The Insides® Neo.
Rate of Refeeding
It is recommended a syringe infusion pump is used to reinfuse the patient’s chyme via The Insides® Neo. This ensures the reinfusion is controlled and comfortable for the patient. Each patient will have their own tolerance for the rate of refeeding. As a general guide:
Monitoring Bowel Movements
Patients will likely have a bowel movement within the first three days of starting chyme reinfusion with The Insides® Neo. If a bowel movement has not happened within three days, urgent assessment is required.
Ostomy Appliance
The Insides® Neo couples with a Hollister™ Newborn Pouch “Pouchkins™” 3778 and Coloplast® “Sensura® Mio Baby”. This ostomy appliance must be changed when it leaks and The Insides® Neo must be changed at the same time. The Insides® Neo can be used for up to four days.
Assessment of Pain and Discomfort
Regular assessment of abdominal distention and discomfort is necessary. This should be completed a minimum of every 4 – 6 hours.
Device Troubleshooting
Please refer to the Frequently Asked Questions for troubleshooting The Insides® Neo
‘Coloplast’, ‘SenSura® Mio ’ are registered trademarks of Coloplast Corporation. The use of the trademarks ‘Coloplast’, ‘SenSura® Mio ‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Coloplast Corporation. We hereby affirm our recognition of Coloplast Corporation 's exclusive intellectual property rights in the trademark ‘Coloplast’, ‘SenSura® Mio,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
‘Hollister’, ‘Pouchkins™’ are registered trademarks of Hollister Incorporated. The use of the trademarks ‘Hollister’, ‘Pouchkins™‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Hollister Incorporated. We hereby affirm our recognition of Hollister Incorporated 's exclusive intellectual property rights in the trademark ‘Hollister’, ‘Pouchkins™,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
Along with normal patient management, the following patient management and assessment is recommended for a patient using The Insides® Neo.
Rate of Refeeding
It is recommended a syringe infusion pump is used to reinfuse the patient’s chyme via The Insides® Neo. This ensures the reinfusion is controlled and comfortable for the patient. Each patient will have their own tolerance for the rate of refeeding. As a general guide:
Monitoring Bowel Movements
Patients will likely have a bowel movement within the first three days of starting chyme reinfusion with The Insides® Neo. If a bowel movement has not happened within three days, urgent assessment is required.
Ostomy Appliance
The Insides® Neo couples with a Hollister™ Newborn Pouch “Pouchkins™” 3778 and Coloplast® “Sensura® Mio Baby”. This ostomy appliance must be changed when it leaks and The Insides® Neo must be changed at the same time. The Insides® Neo can be used for up to four days.
Assessment of Pain and Discomfort
Regular assessment of abdominal distention and discomfort is necessary. This should be completed a minimum of every 4 – 6 hours.
Device Troubleshooting
Please refer to the Frequently Asked Questions for troubleshooting The Insides® Neo
‘Coloplast’, ‘SenSura® Mio ’ are registered trademarks of Coloplast Corporation. The use of the trademarks ‘Coloplast’, ‘SenSura® Mio ‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Coloplast Corporation. We hereby affirm our recognition of Coloplast Corporation 's exclusive intellectual property rights in the trademark ‘Coloplast’, ‘SenSura® Mio,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
‘Hollister’, ‘Pouchkins™’ are registered trademarks of Hollister Incorporated. The use of the trademarks ‘Hollister’, ‘Pouchkins™‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Hollister Incorporated. We hereby affirm our recognition of Hollister Incorporated 's exclusive intellectual property rights in the trademark ‘Hollister’, ‘Pouchkins™,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
Along with normal patient management, the following patient management and assessment is recommended for a patient using The Insides® Neo.
Rate of Refeeding
It is recommended a syringe infusion pump is used to reinfuse the patient’s chyme via The Insides® Neo. This ensures the reinfusion is controlled and comfortable for the patient. Each patient will have their own tolerance for the rate of refeeding. As a general guide:
Monitoring Bowel Movements
Patients will likely have a bowel movement within the first three days of starting chyme reinfusion with The Insides® Neo. If a bowel movement has not happened within three days, urgent assessment is required.
Ostomy Appliance
The Insides® Neo couples with a Hollister™ Newborn Pouch “Pouchkins™” 3778 and Coloplast® “Sensura® Mio Baby”. This ostomy appliance must be changed when it leaks and The Insides® Neo must be changed at the same time. The Insides® Neo can be used for up to four days.
Assessment of Pain and Discomfort
Regular assessment of abdominal distention and discomfort is necessary. This should be completed a minimum of every 4 – 6 hours.
Device Troubleshooting
Please refer to the Frequently Asked Questions for troubleshooting The Insides® Neo
‘Coloplast’, ‘SenSura® Mio ’ are registered trademarks of Coloplast Corporation. The use of the trademarks ‘Coloplast’, ‘SenSura® Mio ‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Coloplast Corporation. We hereby affirm our recognition of Coloplast Corporation 's exclusive intellectual property rights in the trademark ‘Coloplast’, ‘SenSura® Mio,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
‘Hollister’, ‘Pouchkins™’ are registered trademarks of Hollister Incorporated. The use of the trademarks ‘Hollister’, ‘Pouchkins™‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Hollister Incorporated. We hereby affirm our recognition of Hollister Incorporated 's exclusive intellectual property rights in the trademark ‘Hollister’, ‘Pouchkins™,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
Along with normal patient management, the following patient management and assessment is recommended for a patient using The Insides® Neo.
Rate of Refeeding
It is recommended a syringe infusion pump is used to reinfuse the patient’s chyme via The Insides® Neo. This ensures the reinfusion is controlled and comfortable for the patient. Each patient will have their own tolerance for the rate of refeeding. As a general guide:
Monitoring Bowel Movements
Patients will likely have a bowel movement within the first three days of starting chyme reinfusion with The Insides® Neo. If a bowel movement has not happened within three days, urgent assessment is required.
Ostomy Appliance
The Insides® Neo couples with a Hollister™ Newborn Pouch “Pouchkins™” 3778 and Coloplast® “Sensura® Mio Baby”. This ostomy appliance must be changed when it leaks and The Insides® Neo must be changed at the same time. The Insides® Neo can be used for up to four days.
Assessment of Pain and Discomfort
Regular assessment of abdominal distention and discomfort is necessary. This should be completed a minimum of every 4 – 6 hours.
Device Troubleshooting
Please refer to the Frequently Asked Questions for troubleshooting The Insides® Neo
‘Coloplast’, ‘SenSura® Mio ’ are registered trademarks of Coloplast Corporation. The use of the trademarks ‘Coloplast’, ‘SenSura® Mio ‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Coloplast Corporation. We hereby affirm our recognition of Coloplast Corporation 's exclusive intellectual property rights in the trademark ‘Coloplast’, ‘SenSura® Mio,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
‘Hollister’, ‘Pouchkins™’ are registered trademarks of Hollister Incorporated. The use of the trademarks ‘Hollister’, ‘Pouchkins™‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Hollister Incorporated. We hereby affirm our recognition of Hollister Incorporated 's exclusive intellectual property rights in the trademark ‘Hollister’, ‘Pouchkins™,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
Along with normal patient management, the following patient management and assessment is recommended for a patient using The Insides® Neo.
Rate of Refeeding
It is recommended a syringe infusion pump is used to reinfuse the patient’s chyme via The Insides® Neo. This ensures the reinfusion is controlled and comfortable for the patient. Each patient will have their own tolerance for the rate of refeeding. As a general guide:
Monitoring Bowel Movements
Patients will likely have a bowel movement within the first three days of starting chyme reinfusion with The Insides® Neo. If a bowel movement has not happened within three days, urgent assessment is required.
Ostomy Appliance
The Insides® Neo couples with a Hollister™ Newborn Pouch “Pouchkins™” 3778 and Coloplast® “Sensura® Mio Baby”. This ostomy appliance must be changed when it leaks and The Insides® Neo must be changed at the same time. The Insides® Neo can be used for up to four days.
Assessment of Pain and Discomfort
Regular assessment of abdominal distention and discomfort is necessary. This should be completed a minimum of every 4 – 6 hours.
Device Troubleshooting
Please refer to the Frequently Asked Questions for troubleshooting The Insides® Neo
‘Coloplast’, ‘SenSura® Mio ’ are registered trademarks of Coloplast Corporation. The use of the trademarks ‘Coloplast’, ‘SenSura® Mio ‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Coloplast Corporation. We hereby affirm our recognition of Coloplast Corporation 's exclusive intellectual property rights in the trademark ‘Coloplast’, ‘SenSura® Mio,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
‘Hollister’, ‘Pouchkins™’ are registered trademarks of Hollister Incorporated. The use of the trademarks ‘Hollister’, ‘Pouchkins™‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Hollister Incorporated. We hereby affirm our recognition of Hollister Incorporated 's exclusive intellectual property rights in the trademark ‘Hollister’, ‘Pouchkins™,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
Along with normal patient management, the following patient management and assessment is recommended for a patient using The Insides® Neo.
Rate of Refeeding
It is recommended a syringe infusion pump is used to reinfuse the patient’s chyme via The Insides® Neo. This ensures the reinfusion is controlled and comfortable for the patient. Each patient will have their own tolerance for the rate of refeeding. As a general guide:
Monitoring Bowel Movements
Patients will likely have a bowel movement within the first three days of starting chyme reinfusion with The Insides® Neo. If a bowel movement has not happened within three days, urgent assessment is required.
Ostomy Appliance
The Insides® Neo couples with a Hollister™ Newborn Pouch “Pouchkins™” 3778 and Coloplast® “Sensura® Mio Baby”. This ostomy appliance must be changed when it leaks and The Insides® Neo must be changed at the same time. The Insides® Neo can be used for up to four days.
Assessment of Pain and Discomfort
Regular assessment of abdominal distention and discomfort is necessary. This should be completed a minimum of every 4 – 6 hours.
Device Troubleshooting
Please refer to the Frequently Asked Questions for troubleshooting The Insides® Neo
‘Coloplast’, ‘SenSura® Mio ’ are registered trademarks of Coloplast Corporation. The use of the trademarks ‘Coloplast’, ‘SenSura® Mio ‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Coloplast Corporation. We hereby affirm our recognition of Coloplast Corporation 's exclusive intellectual property rights in the trademark ‘Coloplast’, ‘SenSura® Mio,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
‘Hollister’, ‘Pouchkins™’ are registered trademarks of Hollister Incorporated. The use of the trademarks ‘Hollister’, ‘Pouchkins™‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Hollister Incorporated. We hereby affirm our recognition of Hollister Incorporated 's exclusive intellectual property rights in the trademark ‘Hollister’, ‘Pouchkins™,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
Along with normal patient management, the following patient management and assessment is recommended for a patient using The Insides® Neo.
Rate of Refeeding
It is recommended a syringe infusion pump is used to reinfuse the patient’s chyme via The Insides® Neo. This ensures the reinfusion is controlled and comfortable for the patient. Each patient will have their own tolerance for the rate of refeeding. As a general guide:
Monitoring Bowel Movements
Patients will likely have a bowel movement within the first three days of starting chyme reinfusion with The Insides® Neo. If a bowel movement has not happened within three days, urgent assessment is required.
Ostomy Appliance
The Insides® Neo couples with a Hollister™ Newborn Pouch “Pouchkins™” 3778 and Coloplast® “Sensura® Mio Baby”. This ostomy appliance must be changed when it leaks and The Insides® Neo must be changed at the same time. The Insides® Neo can be used for up to four days.
Assessment of Pain and Discomfort
Regular assessment of abdominal distention and discomfort is necessary. This should be completed a minimum of every 4 – 6 hours.
Device Troubleshooting
Please refer to the Frequently Asked Questions for troubleshooting The Insides® Neo
‘Coloplast’, ‘SenSura® Mio ’ are registered trademarks of Coloplast Corporation. The use of the trademarks ‘Coloplast’, ‘SenSura® Mio ‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Coloplast Corporation. We hereby affirm our recognition of Coloplast Corporation 's exclusive intellectual property rights in the trademark ‘Coloplast’, ‘SenSura® Mio,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.
‘Hollister’, ‘Pouchkins™’ are registered trademarks of Hollister Incorporated. The use of the trademarks ‘Hollister’, ‘Pouchkins™‘in these instructions is strictly for informational purposes to provide accurate product information to our customers. The Insides Company Limited does not connote any endorsement, affiliation, sponsorship, or approval by Hollister Incorporated. We hereby affirm our recognition of Hollister Incorporated 's exclusive intellectual property rights in the trademark ‘Hollister’, ‘Pouchkins™,' and any references to said trademark are made in good faith to truthfully describe the compatibility or usage of their product.